- Plunger Over Pressure in Waters HPLC (Alliance System)
Posted by : kaushik zala
Wednesday, March 2, 2011
Today i am working with alliance system and found this problem very frequently, now trying to find practical solutions for this problem......................below mentioned answer is given about problem in waters web site......................
Question:
What does a Plunger Homing Over Pressure error mean?
Answer:
The Plunger Homing Over Pressure error reports an unexpected restriction in the flow path during initialization of the piston drives. The most likely items to check would be the pump's inline filter or precolumn filter.
yours chromatographically,
kaushik zala
taken from waters website with thanks...........................
practical solution is that , i have changed column inlet tubing and now system is working with optimum pressure range................