- Back to Home »
- PROBLEMS WITH CHROMATOGRAM
Posted by : kaushik zala
Wednesday, December 14, 2011
Many problems in an LC system show up as changes in the chromatogram. Some of these can be solved by changes in the equipment; however, others require modification of the assay procedure. Selecting the proper column type and mobile phase are keys to "good chromatography."
A. Peak Tailing
Possible
Cause
|
Solution
|
|
|
1.
Blocked frit
|
1.
a. Reverse flush column (if allowed)
|
|
|
|
|
2.
Column void
|
2.
Fill void
|
3.
Interfering peak
|
3.
a. Use longer column
|
|
|
|
|
4.
Wrong mobile-phase pH
|
4.
a. Adjust pH
|
|
|
|
|
5.
Sample reacting with active sites
|
5.
a. Add ion pair reagent or volatile basic
|
|
|
B. Peak Fronting
Possible
Cause
|
Solution
|
|
|
1.
Low temperature
|
1.
Increase column temperature
|
2.
Wrong sample solvent
|
2.
Use mobile phase for injection solvent
|
3.
Sample overload
|
3.
Decrease sample concentration
|
4.
Bad column
|
C. Split Peaks
Possible
Cause
|
Solution
|
|
|
1.
Contamination on guard or
|
1.
a. Remove guard column and attempt
|
|
|
|
|
2.
Sample solvent incompatible with
|
2.
Change solvent; whenever possible,
|
|
|
D. Distortion of
Larger Peaks
Possible
Cause
|
Solution
|
|
|
1.
Sample overload
|
1.
Reduce sample size
|
E. Distortion of Early
Peaks
Possible
Cause
|
Solution
|
|
|
1.
Wrong injection solvent
|
1.
a. Reduce injection volume
|
|
|
F. Tailing, Early
Peaks More Than Later Ones
Possible
Cause
|
Solution
|
|
|
1.
Extra-column effects
|
1.
a. Replumb system (shorter, narrower
|
|
|
|
|
G. Increased Tailing as
k' Increases
Possible
Cause
|
Solution
|
|
|
1.
Secondary retention effects,
|
1.
a. Add triethylamine (basic samples)
|
|
|
|
|
|
|
2.
Secondary retention effects,
|
2.
a. Add triethylamine (basic compounds)
|
|
|
3.
Secondary retention effects,
|
3.
Add triethylamine (basic samples)
|
|
|
H. Acidic or Basic
Peaks Tail
Possible
Cause
|
Solution
|
|
|
1.
Inadequate buffering
|
1.
a. Use 50–100 mM buffer concentration
|
|
|
|
|
I. Extra Peaks
Possible
Cause
|
Solution
|
|
|
1.
Other components in sample
|
1.
Normal
|
2.
Late-eluting peak from previous
|
2.
a. Increase run time or gradient slope
|
|
|
3.
Vacancy or ghost peaks
|
3.
a. Check purity of mobile phase
|
|
|
|
|
J. Retention Time
Drifts
Possible
Cause
|
Solution
|
|
|
1.
Poor temperature control
|
1.
Thermostat column
|
2.
Mobile phase changing
|
2.
Prevent change (evaporation, reaction,
|
|
|
3.
Poor column equilibration
|
3.
Allow more time for column equili-
|
|
|
K. Abrupt Retention
Time Changes
Possible
Cause
|
Solution
|
|
|
1.
Flow rate change
|
1.
Reset flow rate
|
2.
Air bubble in pump
|
2.
Bleed air from pump
|
3.
Improper mobile phase
|
3.
a. Replace with proper mobile phase
|
|
|
|
|
L. Baseline Drift
Possible
Cause
|
Solution
|
|
|
1.
Column temperature fluctuation
|
1.
a. Control column and mobile phase
|
|
|
|
|
|
|
|
|
|
|
|
|
2.
Nonhomogeneous mobile phase
|
2.
a. Use HPLC-grade solvents, high-purity
|
|
|
|
|
|
|
|
|
3.
Contaminant or air buildup in
|
3.
a. Flush cell with methanol or other
|
|
|
|
|
|
|
4.
Plugged outlet line after detector
|
4.
a. Unplug or replace line
|
|
|
|
|
5.
Mobile-phase mixing problem or
|
5.
a. Correct composition/flow rate
|
|
|
|
|
6.
Slow column equilibration,
|
6.
a. Flush with intermediate strength
|
|
|
|
|
|
|
7.
Mobile phase contaminated,
|
7.
a. Check make-up of mobile phase
|
|
|
|
|
8.
Strongly retained materials in
|
8.
a. Use guard column
|
|
|
|
|
|
|
9.
Mobile phase recycled but
|
9.
a. Reset baseline
|
|
|
|
|
10.
Detector (UV) not set at
|
10.
Change wavelength to UV absorbance
|
|
|
|
|
|
M. Baseline Noise
(Regular)
Possible
Cause
|
Solution
|
|
|
1.
Air in mobile phase, detector cell,
|
1.
a. Degas mobile phase
|
|
|
|
|
2.
Leak
|
|
|
|
3.
Incomplete mobile phase mixing
|
3.
Mix mobile phase by hand or use less
|
|
|
4.
Temperature effect (column at
|
4.
Reduce differential or add head
|
|
|
|
|
5.
Other electronic equipment on
|
5.
Isolate LC, detector, or recorder to
|
|
|
|
|
6.
Pump pulsations
|
6.
Incorporate pulse dampener into system
|
N. Baseline Noise
(Irregular)
Possible
Cause
|
Solution
|
|
|
1.
Leak
|
|
|
|
2.
Mobile phase contaminated,
|
2.
Check make-up of mobile phase
|
|
|
|
|
3.
Mobile phase solvents immiscible
|
3.
Select and use only miscible solvents
|
4.
Detector/recorder electronics
|
4.
a. Isolate detector and recorder
|
|
|
|
|
|
|
5.
Air trapped in system
|
5.
Flush system with strong solvent
|
6.
Air bubbles in detector
|
6.
a. Purge detector
|
|
|
|
|
7.
Detector cell contaminated (even
|
7.
Clean cell by flushing with 1 NHNO3
|
|
|
|
|
8.
Weak detector lamp
|
8.
Replace lamp
|
9.
Column leaking silica or packing
|
9.
Replace column
|
|
|
10.
Mobile phase mixer inadequate or
|
10.
Repair or replace the mixer or mix off-
|
|
|
O. Broad Peaks
Possible
Cause
|
Solution
|
|
|
1.
Mobile-phase composition changed
|
1.
Prepare new mobile phase
|
2.
Mobile-phase flow rate too low
|
2.
Adjust flow rate
|
3.
Leaks (especially between column
|
|
|
|
|
|
|
|
|
|
4.
Detector settings incorrect
|
4.
Adjust settings
|
5.
Extra-column effects:
|
5.
a. Inject smaller column (e.g., 10 µl vs.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
6.
Buffer concentration too low
|
6.
Increase concentration
|
7.
Guard column contaminated/worn
|
7.
Replace guard column
|
|
|
8.
Column contaminated/worn out;
|
8.
a. Replace column with new one of same
|
|
|
|
|
|
|
|
|
9.
Void at column inlet
|
9.
Open inlet end and fill void or replace
|
|
|
10.
Peak represents two or more
|
10.
Change column type to improve
|
|
|
11.
Column temperature too low
|
11.
Increase temperature; do not exceed
|
|
|
|
|
|
|
12.
Detector time constant too large
|
12.
Use smaller time constant
|
|
P. Loss of Resolution
Possible
Cause
|
Solution
|
|
|
1.
Mobile phase contaminated/
|
1.
Prepare new mobile phase
|
|
|
|
|
2.
Obstructed guard or analytical
|
2.
a. Remove guard column and attempt
|
|
|
|
|
Q. All Peaks Too Small
Possible
Cause
|
Solution
|
|
|
1.
Detector attentuation too high
|
1.
Reduce attenuation
|
2.
Detector time constant too large
|
2.
Use smaller time constant
|
3.
Injection size too small
|
3.
Use larger sample loop
|
4.
Improper recorder connection
|
4.
Use correct connection
|
R. All Peaks Too large
Possible
Cause
|
Solution
|
|
|
1.
Detector attenuation too low
|
1.
Use larger attenuation
|
2.
Injection size too large
|
2.
Use smaller sample loop
|
3.
Improper recorder connection
|
3.
Use correct connection
copied
from
|
yours
chromatogaraphically,
KAUSHIK ZALA