Posted by : kaushik zala Wednesday, December 14, 2011


Many problems in an LC system show up as changes in the chromatogram. Some of these can be solved by changes in the equipment; however, others require modification of the assay procedure. Selecting the proper column type and mobile phase are keys to "good chromatography."
A. Peak Tailing
Possible Cause
Solution

1. Blocked frit
1. a. Reverse flush column (if allowed)

b. Replace inlet frit

c. Replace column
2. Column void
2. Fill void
3. Interfering peak
3. a. Use longer column

b. Change mobile-phase and/or column/

selectivity
4. Wrong mobile-phase pH
4. a. Adjust pH

b. For basic compounds, a lower pH

usually provides more symmetric peaks
5. Sample reacting with active sites
5. a. Add ion pair reagent or volatile basic

modifier
b. Change column
B. Peak Fronting
Possible Cause
Solution

1. Low temperature
1. Increase column temperature
2. Wrong sample solvent
2. Use mobile phase for injection solvent
3. Sample overload
3. Decrease sample concentration
4. Bad column
4. See A.1. and A.2.
C. Split Peaks
Possible Cause
Solution

1. Contamination on guard or
1. a. Remove guard column and attempt
analytical column inlet
analysis

b. Replace guard if necessary
c. If analytical column is obstructed,
reverse and flush
d. If problem persists, column may be
fouled with strongly retained
contaminants
e. Use appropriate restoration procedure
f. If problem persists, inlet is probably
plugged
g. Change frit or replace column
2. Sample solvent incompatible with
2. Change solvent; whenever possible,
mobile phase
inject samples in mobile phase
D. Distortion of Larger Peaks
Possible Cause
Solution

1. Sample overload
1. Reduce sample size
E. Distortion of Early Peaks
Possible Cause
Solution

1. Wrong injection solvent
1. a. Reduce injection volume

b. Use weaker injection solvent
F. Tailing, Early Peaks More Than Later Ones
Possible Cause
Solution

1. Extra-column effects
1. a. Replumb system (shorter, narrower

tubing)

b. Use smaller volume detector cell
G. Increased Tailing as k' Increases
Possible Cause
Solution

1. Secondary retention effects,
1. a. Add triethylamine (basic samples)
reversed-phase mode
b. Add acetate (acidic samples)

c. Add salt or buffer (ionic samples)

d. Try a different column
2. Secondary retention effects,
2. a. Add triethylamine (basic compounds)
normal-phase mode
b. Add acetic acid
3. Secondary retention effects,
3. Add triethylamine (basic samples)
ion-pair

H. Acidic or Basic Peaks Tail
Possible Cause
Solution

1. Inadequate buffering
1. a. Use 50–100 mM buffer concentration

b. Use buffer with pKa equal to pH of

mobile phase
I. Extra Peaks
Possible Cause
Solution

1. Other components in sample
1. Normal
2. Late-eluting peak from previous
2. a. Increase run time or gradient slope
injection
b. Increase flow rate
3. Vacancy or ghost peaks
3. a. Check purity of mobile phase

b. Use mobile phase as injection solvent

c. Reduce injection volume
J. Retention Time Drifts
Possible Cause
Solution

1. Poor temperature control
1. Thermostat column
2. Mobile phase changing
2. Prevent change (evaporation, reaction,

etc.)
3. Poor column equilibration
3. Allow more time for column equili-

bration between runs
K. Abrupt Retention Time Changes
Possible Cause
Solution

1. Flow rate change
1. Reset flow rate
2. Air bubble in pump
2. Bleed air from pump
3. Improper mobile phase
3. a. Replace with proper mobile phase

b. Set proper mobile phase mixture on

controller
L. Baseline Drift
Possible Cause
Solution

1. Column temperature fluctuation
1. a. Control column and mobile phase
(even small changes cause cyclic
temperature
baseline rise and fall; most
b. Use heat exchanger before detector
often affects refractive index

and conductivity detectors, or

UV detectors at high sensitivity

or in direct photometric mode)

2. Nonhomogeneous mobile phase
2. a. Use HPLC-grade solvents, high-purity
(drift usually to higher
salts, and additives
absorbance, rather than cyclic
b. Degas mobile phase before use
pattern from temperature
c. Sparge with helium during use
fluctuation)

3. Contaminant or air buildup in
3. a. Flush cell with methanol or other
detector cell
strong solvent

b. If necessary, clean cell with 1 N

HNO3 (never with HCl)
4. Plugged outlet line after detector
4. a. Unplug or replace line
(high-pressure cracks cell
b. Refer to detector manual to replace
window, producing noisy baseline)
window
5. Mobile-phase mixing problem or
5. a. Correct composition/flow rate
change in flow rate
b. To avoid, routinely monitor

composition and flow rate
6. Slow column equilibration,
6. a. Flush with intermediate strength
especially when changing
solvent
mobile phase
b. Run 10–20 column volumes of new

mobile phase before analysis
7. Mobile phase contaminated,
7. a. Check make-up of mobile phase
deteriorated, or prepared from
b. Use highest grade chemicals and HPLC
low-quality materials
solvents
8. Strongly retained materials in
8. a. Use guard column
sample (high k) can elute as very
b. If necessary, flush column with strong
broad peaks and appear to be
solvent between injections or periodi-
a problem
ically during analysis
9. Mobile phase recycled but
9. a. Reset baseline
detector not adjusted
b. Use new mobile phase when dynamic

range of detector is exceeded
10. Detector (UV) not set at
10. Change wavelength to UV absorbance
absorbance maximum but at slope
maximum
of curve


M. Baseline Noise (Regular)
Possible Cause
Solution

1. Air in mobile phase, detector cell,
1. a. Degas mobile phase
or pump
b. Flush system to remove air from

detector cell or pump
2. Leak
2. a. See Section 3

b. Check system for loose fittings
c. Check pump for leaks, salt build-up,
unusual noises
d. Change pump seals if necessary
3. Incomplete mobile phase mixing
3. Mix mobile phase by hand or use less

viscous solvent
4. Temperature effect (column at
4. Reduce differential or add head
high temperature, detector
exchanger
unheated)

5. Other electronic equipment on
5. Isolate LC, detector, or recorder to
same line
determine if source of problem is

external; correct as necessary
6. Pump pulsations
6. Incorporate pulse dampener into system
N. Baseline Noise (Irregular)
Possible Cause
Solution

1. Leak
1. a. See Section 3

b. Check for loose fittings
c. Check pump for leaks, salt build-up,
unusual noises
d. Change seals if necessary
e. Check for detector cell leak
2. Mobile phase contaminated,
2. Check make-up of mobile phase
deteriorated, or prepared

from low-quality materials

3. Mobile phase solvents immiscible
3. Select and use only miscible solvents
4. Detector/recorder electronics
4. a. Isolate detector and recorder

electronically

b. Refer to instruction manual to correct

problem
5. Air trapped in system
5. Flush system with strong solvent
6. Air bubbles in detector
6. a. Purge detector

b. Install back-pressure device after

detector
7. Detector cell contaminated (even
7. Clean cell by flushing with 1 NHNO3
small amounts of contaminants
(never with HCl)
can cause noise)

8. Weak detector lamp
8. Replace lamp
9. Column leaking silica or packing
9. Replace column
material

10. Mobile phase mixer inadequate or
10. Repair or replace the mixer or mix off-
malfunctioning
line if isocratic
O. Broad Peaks
Possible Cause
Solution

1. Mobile-phase composition changed
1. Prepare new mobile phase
2. Mobile-phase flow rate too low
2. Adjust flow rate
3. Leaks (especially between column
3. a. See Section 3
and detector)
b. Check for loose fittings

c. Check pump for leaks, salt build-up,

and unusual noises

d. Change seals if necessary
4. Detector settings incorrect
4. Adjust settings
5. Extra-column effects:
5. a. Inject smaller column (e.g., 10 µl vs.
a. Column overloaded
100 µl) or 1:10 and 1:100 dilutions of
b. Detector response time or cell
sample
volume too large
b. Reduce response time or use smaller

cell
c. Tubing between column and
c. Use as short a piece of 0.007–0.010-
detector too long or ID
inch ID tubing as practical
too large

d. Recorder response time too
d. Reduce response time
high

6. Buffer concentration too low
6. Increase concentration
7. Guard column contaminated/worn
7. Replace guard column
out

8. Column contaminated/worn out;
8. a. Replace column with new one of same
low plate number
type

b. If new column provides symmetrical

peaks, flush old column with strong

solvent
9. Void at column inlet
9. Open inlet end and fill void or replace

column
10. Peak represents two or more
10. Change column type to improve
poorly resolved compounds
separation
11. Column temperature too low
11. Increase temperature; do not exceed

75°C unless higher temperatures

are acceptable to column

manufacturer
12. Detector time constant too large
12. Use smaller time constant

P. Loss of Resolution
Possible Cause
Solution

1. Mobile phase contaminated/
1. Prepare new mobile phase
deteriorated (causing

retention time to change)

2. Obstructed guard or analytical
2. a. Remove guard column and attempt
column
analysis

b. Replace guard if necessary
c. If analytical column is obstructed,
reverse and flush; if problem
persists, column may be fouled
with strongly retained contaminants
d. Use appropriate restoration procedure;
if problem persists, inlet is probably
plugged
e. Change frit or replace column
Q. All Peaks Too Small
Possible Cause
Solution

1. Detector attentuation too high
1. Reduce attenuation
2. Detector time constant too large
2. Use smaller time constant
3. Injection size too small
3. Use larger sample loop
4. Improper recorder connection
4. Use correct connection
R. All Peaks Too large
Possible Cause
Solution

1. Detector attenuation too low
1. Use larger attenuation
2. Injection size too large
2. Use smaller sample loop
3. Improper recorder connection
3. Use correct connection



copied from 

yours chromatogaraphically,

KAUSHIK ZALA

Leave a Reply

Subscribe to Posts | Subscribe to Comments

About Me

My photo
gandhinagar, gujarat, India
HPLC LOVER

Labels

MY PHOTOS ON FACE BOOK

MY OLDER BLOGS

MY FACE BOOK PROFILE

KAUSHIK ZALA, HPLC SCIENTIST. Powered by Blogger.
Copyright © 2013 Yours Chromatographically, Kaushik Zala | Dark Simple Blogger Template Powered by Blogger | Created by Renadel Dapize | Ori. BRS-bt Djogzs | All Rights Reserved