Posted by : kaushik zala Thursday, March 1, 2012


Start up - Preliminary checks

ProblemPossible causeSolution
No peaks or very small peaksDetector offCheck detector
Broken connections to recorderCheck connections
No sample/Wrong sampleCheck sample. Be sure it is not deteriorated. Check for bubbles in the vials
Wrong settings on recorder or detectorCheck attenuation. Check gain
No FlowPump offStart Pump
Flow interruptedCheck reservoirs. Check position of the inlet tubing. Check loop for obstruction or air. Check degasing of mobile phase. Check compatibility of the mobile phase components.
LeakCheck fittings. Check pump for leaks and precipitates. Check pump seals.
Air trapped in the systemDisconnect column and prime pump. Flush system with 100% methanol or isopropanol. Contact servicing if necessary.

Column and Fittings Leaks

ProblemPossible causeSolution
Column end leaksLoose fitting
White powder at loose fitting
Tighten or replace fitting
Cut tubing and replace ferrule; disassemble fitting, rinse and reassemble.
Leak at detectorDetector-seal failureReplace detector seal or gaskets.
Leak at injection valveWorn or scratched valve rotorReplace valve rotor
Leak at pumpPump seal failureReplace pump seal; check piston for scratches and, if necessary, replace

Change in Retention time

ProblemPossible causeSolution
Changing Retention TimesBuffer retention timesUse buffer with concentration greater than 20 mM.
Contamination buildupFlush column occasionally with strong solvent
Equilibration time insufficient for gradient run or changes in isocratic mobile phasePass at least 10 column volumes through the column for gradient regeneration or after solvent changes
First few injections - active sitesCondition column by injecting concentrated sample
Inconsistent on-line mobile-phase mixingEnsure gradient system is delivering a constant composition; compare with manually prepared mobile phase; partially premix mobile phase
Selective evaporation of mobile-phase componentCover solvent reservoirs; use less-vigorous helium purging; prepare fresh mobile phase
Varying column temperatureThermostat or insulate column; ensure laboratory temperature is constant.
Decreasing Retention TimesActive sites on column packingUse mobil-phase modifier, competing base (basic compounds), or increase buffer strength; use higher coverage column packing.
Column overloaded with sampleDecrease sample amount or use larger-diameter column.
Increasing flow rateCheck and reset pump flow rate.
Loss of bonded stationary phase or base silicaUse mobile-phase pH between pH 2 and pH 8
Varying column temperatureThermostat or insulate column; ensure laboratory temperature is constant
Increasing Retention TimesDecreasing flow rateCheck and reset pump flow rate; check for pump cavitation; check for leaking pump seals and other leaks in system
Changing mobile-phase compositionCover solvent reservoirs; ensure that gradient system is delivering correct composition.
Loss of bonded stationary phaseUse mobile-phase pH between pH 2 and pH 8
Slow column equilibration timeReversed phase ion pairing - long chain ion pairing reagents require longer equilibration timeUse ion-pairing reagent with shorter alkyl chain length

Baseline

ProblemPossible causeSolution
Void Time noiseAir bubbles in mobile phaseDegas or use back pressure restricor on detector
Positive-negative - difference in refractive index of injection solvent and mobile phaseNormal with many samples; use mobile phase as sample solvent
Drifting baselineNegative direction (gradient elution) - absorbance of mobile-phase AUse non-UV absorbing mobile phase solvents; use HPLC grade mobile phase solvents; add UV absorbing compound to mobile phase B.
Positive direction (gradient elution) - absorbance of mobile phase BUse higher UV absorbance detector wavelength; use non-UV absorbing mobile phase solvents; use HPLC grade mobile phase solvents; add UV absorbing compound to modile phase A.
Positive direction - contamination buildup and elutionFlush column with strong solvent; clean up sample; use HPLC grade solvents
Wavy or undulating - temperature changes in roomMonitor and control changes in room temperature; insulate column or use column oven; cover refractive index detector and keep it out of air currents.
Baseline noiseContinous - detector lamp problem or dirty cellReplace UV lamp( each should last 2000 h; clean and flush flow cell.
Gradient or isocratic proportioning - lack of solvent mixingUse proper mixing device; check proportioning precision by spiking one solvent with UV absorbing compound and mointor UV absorbance detector outputl.
Gradient or isocratic proportioning - malfunctioning proportioning valveslClean or replace proportioning precision valves; partially remix solventsl.
Occasional sharp spikes - external electrical interferenceUse voltage stabilizer for LC system; use independent electrical circuit.
Periodic - pump pulsesService or replace pulse damper; purge air from pump; clean or replace check valves.
Random - contamination buildupFlush column with strong solvent; clean up sample; use HPLC grade solvent
Spikes - bubble in detectorDegas mobile phase; use back pressure restrictor at detector outlet.
Spikes - column temperature higher than boiling point of solventUse lower column temperature.

Pressure

ProblemPossible causeSolution
Decreasing PressureInsufficient flow from pumpLoosen cap on mobile phase reservior
Leak in hydralic lines from pump to columnTighten or replace fittings; tighten rotor in injection valve
Leaking pump check valve or sealsReplace or clean check valves; replace pump seals.
Pump cavitationDegas solvent; check for obstruction in line from solvent reservoir to pump; replace inlet-line frit
Fluctuating pressurreBubble in pumpDegas solvent; purge solvent with helium
Leaking pump check valve or sealsReplace or clean check valves; replace pump seals
High Back PressureColumn blocked wth irreversibly adorbed sampleImprove sample cleanup; use guard column; reverse-flush column with strong solvent to dissolve blockage
Column particle size too small (for example 3 micrometers)Use larger particle size (for example 5 micrometer)
Microbial growth on columnUse at least 10% organic modifier in mobile phase; use fresh buffer daily; add 0.02% sodium azide to aqueous mobile phase; store column in at least 25% organic solvent without buffer
Mobile phase viscosity too highUse lower viscosity solvents or higher temperature
Plugged frit in in-line filter or guard columnReplace frit or guard column
Plugged inlet fritReplace endfitting or frit assembly
Polymetric columns - solvent change causes swelling of packingUse correct solvent with column; change to proper solvent compositionl consult manufacturer's solvent-compatibility chartl use a column with a higher percentage of cross-linking
Salt precipitation (especially in reversed-phase chromatography with high concentration of organic solvent in mobile phase) concentration of organic solvent in mobile phase)Ensure mobile phase compatibility with buffer concentration; decrease ionic strength and water-organic solvent ratio; premix mobile phase
When injector disconnected from column - blockage in injectorClean injector or replace rotor
Increasing PressureBlocked flow linesSystematically disconnect components from detector end to column end to find blockage; replace or clean blocked component
Particulate buildup at head of columnFilter sample; use .5 micrometer in-line filter; disconnect and backflush column; replace inlet frit
Water-organic solvent systems - buffer precipitationEnsure mobile phase compatibility with buffer concentration; decrease ionic strength or water organic solvent ratio

Peaks

ProblemPossible causeSolution
Broad peaksAnalytes eluted early due to sample overloadDilute sample 1:10 and reinject
Detector-cell volume too largeUse smallest possible cell volume consistent with sensitivity needs; use detector with no heat exchanger in system
Injection volume too largeDecrease solvent strength of injection solvent to focus solute; inject smaller volume
Large extra column volumeUse low- or zero-dead-volume endfittings and connectors; use smallest possible diameter of connecting tubing (<0.10 in. i.d.); connect tubing with matched fittings
Mobile-phase solvent viscosity too highIncrease column temperature; change to lower viscosity solvent
Peak dispersion in injector valveDecrease injector sample loop size; introduce air bubble in front and back of sample in loop
Poor column efficiencyUse smaller-particle-diameter packing, lower-viscosity mobile phase, higher column temperature, or lower flow rate
Retention time too longUse gradient elution or stronger isocratic mobile phase
Sampling rate of data system too lowIncrease sampling frequency.
Slow detector time constantAdjust time constant to match peak width
Some peaks broad - late elution of analytes retained from previous injectionFlush column with strong solvent at end of run; end gradient at higher solvent concentration
Ghost peaksContaminationFlush column to remove contaminatint; use HPLC-grade solven
Elution of analytes retained from previous injectionFlush column with strong solvent at end of run; end gradient at higher solvent concentration
Ion-pair chromatography - upset equilibriumPrepare sample in mobile phase; reduce injection volume
Oxidation of trifluoroacetic acid in peptide mappingPrepare trifluoroacetic acid solutions fresh daily; use antioxidant
Reversed-phase chromatography - contaminated waterCheck suitability of water by running different amounts through column and measure peak height of interferences as function of enrichment time; clean water by running it through old reversed-phase column; use HPLC-grade water.
Unknown interferences in sampleUse sample cleanup or prefractionation before injection.
Negative peaksRefractive index detection - refractive index of solute less than that of mobile phaseReverse polarity to make peak positive
UV-absorbance detection - absorbance of solute less than that of mobile phaseUse mobile phase with lower UV absorbance; if recycling solvent, stop recycling when recycled solvent affects detection
Peak DoublingBlocked FritReplace or clean frit; install 0.5-um porosity in-line filter between pump and injector to eliminate mobile-phase contaminants or between injector and column to eliminate sample contaminants
Coelution of interfering compoundUse sample cleanup or prefractionation; adjust selectivity by changing mobile or stationary phase
Coelution of interfering compound from previous injectionFlush column with strong solvent at end of ran; end gradient at higher solvent concentration
Column overloadedUse higher-capacity stationary phase; increase column diameter; decrease sample amount
Column void or channelingReplace column, or, if possible, open top endfitting and clean and fill void with glass beads or same column packing; repack column
Injection solvent too strongUse weaker injection solvent or stronger mobile phase
Sample volume too largeUse injection volume equal to one-sixth of column volume when sample prepared in mobile phase for injection
Unswept injector flow pathReplace injector rotor
Peak FrontingChanneling in columnReplace or repack column
Column overloadedUse higher-capacity stationary phase; increase column diameter; decrease sample amount
Tailing PeaksBasic solutes - silanol interactionsUse competing base such as triethylamine; use a stronger mobile phase; use base-deactivated silica-based reversed-phase column; use polymeric column
Beginning of peak doublingSee peak doubling
Chelating solutes - trace metals in base silicaUse high purity silica-based column with low trace-metal content; add EDTA or chelating compound to mobile phase; use polymeric column
Silica-based column - degradation at high pHUse polymeric, sterically protected, or high-coverage reversed-phase column; install silica gel saturatorcolumn between pump and injector
Silica-based column - degradation at high temperatureReduce temperature to less than 50 C
Silica-based column - silanol interactionsDecrease mobile-phase pH to suppress silanol ionization; increase buffer concentration; derivatize solute to change polar interactions
Unswept dead volumeMinimize number of connections; ensure injector rotor seal is tight; ensure all compression fittings arecorrectly seated
Void formation at head of columnReplace column, or, if possible, open top endfitting and clean and fill in void with glass beads or samecolumn packing; rotate injection valve quickly; use injection valve with pressure bypass; avoid pressure shock
SpikesBubbles in mobile phaseDegas mobile phase; use back-pressure restrictor at detector outlet; ensure that all fittings are tight
Column stored without capsStore column tightly capped; flush reversed-phase columns with degassed methanol

yours chromatographically.....,


kaushik zala

1 comments

  1. I do have a small understanding on this side of the work, good i have good people surrounding me inside the lab.
    uv detector

    ReplyDelete

About Me

My photo
gandhinagar, gujarat, India
HPLC LOVER

Labels

MY PHOTOS ON FACE BOOK

MY OLDER BLOGS

MY FACE BOOK PROFILE

KAUSHIK ZALA, HPLC SCIENTIST. Powered by Blogger.
Copyright © 2013 Yours Chromatographically, Kaushik Zala | Dark Simple Blogger Template Powered by Blogger | Created by Renadel Dapize | Ori. BRS-bt Djogzs | All Rights Reserved